|
ASHI Governance
Standards for Histocompatibility Testing Section Q - Flow Cytometry These
standards apply to histocompatibility testing and leukocyte
phenotyping by flow cytometry. Q1.000
Instrument Standardization/Calibration Q1.100
An optical standard , consisting of latex beads or other uniform
particles, shall be run to insure proper focusing and alignment of all
lenses in the path for both the exciting light source and signal
(light scatter, fluorescence, etc.) detectors. Q1.110
The optical standard shall be run each time the instrument is turned
on and any time maintenance, adjustments or sample problems likely to
have altered optical alignment (obstruction of fluidics) occur during
operation. Q1.120
The results of optical focusing/alignment must be recorded in a daily
quality control log. Q1.130
A threshold value for acceptable optical standardization must be
established for all relevant signals for each instrument and the
focusing procedure repeated until these values are achieved or
surpassed. Q1.140
In the event a particular threshold value cannot be attained, a
written protocol for instituting corrective action must be available.
This protocol should include appropriate corrective actions including
clear guidelines describing when a service call is warranted. Q1.200
A fluorescent standard for each fluorochrome to be used, shall be run
to insure adequate amplification of the fluorescent signal(s) on a
day-to-day basis. Q1.210
This standard may be incorporated in the beads or other particles used
for optical standardization or may be a separate bead or fixed cell
preparation. Q1.220
The fluorescent standard must be run each time the instrument is
turned on and any time maintenance, adjustments or sample problems
likely to have altered the gain or high voltage settings (e.g.obstruction
of fluidics) occur during operation. Q1.230
The results of fluorescent standardization shall be recorded in a
daily quality control log. Q1.240
In the event that acceptable fluorescence separation cannot be
attained, a written protocol for instituting corrective action must be
available. This protocol should include appropriate corrective action
including clear guidelines describing when a service call is
warranted. Q1.300
If performing analyses that require the simultaneous use of two or
more fluorochromes, an appropriate procedure must be used to
compensate for “spill over” into the other fluorescence detectors. Q1.400
For laser based instruments, the current input (amps) and laser light
output (milliwatts), at the normal operating wavelength measured after
the laser is peaked and normal operating power set, must be recorded
as part of a daily quality control record. Q2.000
FLOW CYTOMETRIC CROSSMATCH TECHNIQUE Q2.100
A multi-color technique is highly recommended. However, if a single
color technique is used, the purity of the isolated cell population
must be documented and should be of sufficient purity to define the
population for analysis. Q2.110
The binding of human immunoglobulin should be assessed with a
fluorochrome labelled (e.g.fluorescein) F(ab’)2 anti-human IgG. Q2.120
Binding of antibody to T cells, B cells and/or monocytes should be
positively confirmed with a differently labelled (e.g. phycoerythrin)
monoclonal antibody that detects the corresponding cluster designated
antigen (e.g.CD3 for T cells, CD19 or CD20 for B cells and CD14 for
monocytes). Q2.130
Multicolor staining of other immunoglobulin classes and target cells
may also be justified. Q2.140
Each laboratory should establish and document the optimum serum/cell
ratio i.e. a standard number of cells to a fixed volume of serum. Q2.200
Controls Q2.210
The normal human serum control should be from a non-alloimmunized and
otherwise healthy individual and must be screened by flow cytometry to
insure lack of reactivity against human lymphocytes. Q2.220
The positive control should be human serum containing antibodies of
the appropriate isotype, specific for the HLA antigens or any other
alloantigens deemed to be important for detection in the crossmatch..
Positive controls should react with lymphocytes of all humans.. Q2.230
The anti-human immunoglobulin reagent should be titered to determine
the dilution with optimal activity (signal to noise ratio). If a
multicolor technique is employed, the reagent must not demonstrate
crossreactivity with the other immunoglobulin reagents used to mark
the cells.. Q2.240
Regardless of the method used for reporting raw data(mean, median,
mode channel shifts or quantitative fluorescence measurements), each
lab must establish its own threshold for discriminating positive
reactions. Any significant change in protocol, reagents or
instrumentation requires repeat determination of the positive
threshold. Q2.300
Interpretation Q2.310
Each laboratory must define the criteria used to define positive and
negative crossmatches. Q3.000
IMMUNOPHENOTYPING BY FLOW CYTOMETRY Q3.100
Terminology used must be defined and/or conform to nomenclature
recommended/approved by the most recent International Workshop of
Differentiation Antigens of Human Leucocytes or other appropriate
scientific organizations. Q3.200
Cell Preparation. Q3.210
The method used for cell preparation should be documented to yield
appropriate preparations of viable cells. Q3.220
The viability of cell preparations should be recorded and should
exceed the laboratory’s established minimum standards for each
procedure used. Q3.230
For internal labelling, the method used to allow fluorochrome labelled
antibodies to penetrate the cell membrane must be documented to be
effective. Q3.300
Labelling of specimens. Q3.310
Specificity controls, consisting of appropriate cell types known to be
positive for selected standard antibodies must be run within
laboratory-defined intervals sufficiently short to assure the proper
performance of reagents. Q3.320
A negative reagent control(s) shall be run for each test cell
preparation. This control should consist of monoclonal antibody(ies)
of the same species and subclass and should be prepared/purified in
the same way as the monoclonal(s) used for phenotyping. Q3.330
For indirect labelling, the negative control reagent should be an
irrelevant primary antibody, if available, and in all cases, the same
secondary antibody(ies) conjugated with the same fluorochrome(s) used
in all relevant test combinations. Q3.340
For direct labelling, the negative control reagent should be an
irrelevant antibody conjugated with the same fluorochrome and at the
same fluorochrome:protein ratio used in all relevant test
combinations. Q3.350
Whether analyzed directly or fixed prior to analysis, labelled cells
must be analyzed within a time period demonstrated by the laboratory
to avoid significant loss of any cell subpopulation or total cell
numbers. Control samples must be analyzed within the same period after
staining as the test samples. Q3.360
If analysis will be based on a population of cells selected by flow
cytometry “gating” on size or density parameters, or selected by
depletion or enrichment techniques, control stains must be run for
each test individual to detect the presence of contaminating cells in
the selected population. (e.g. Monocyte contamination of
‘lymphocytes’ gated by forward angle or forward angle vs 90°
light scatter must be detected with a monocyte specific marker
antibody. Q3.370
Conclusions about abnormal proportions or abnormal numbers of cells
bearing particular internal or cell surface markers must only be drawn
in comparison with local `control’ data obtained with the same
instrument, reagents and techniques. Q3.380
Determination of percent positives must take into consideration the
results of the negative control reagent. However, when clearly defined
positive and negative populations are evident in the test sample, it
may be appropriate to adjust the threshold based on the test sample. Q3.400
REAGENTS Q3.410
The specificity of monoclonal antibodies shall be verified by
published and/or manufacturer’s documentation and whenever possible
verified locally through tests with appropriate control cells prepared
and tested by the same method employed in the laboratory’s test
sample analysis. Q3.420
The quantities of reagents used for each test sample must be
determined by the manufacturers or from published data and whenever
possible should be verified locally by appropriate titration
procedures. Q3.430
Reagents must be stored according to manufacturers’ instructions or
according to conditions verified to maintain stability by documented
local tests. Q3.440
Monoclonal antibodies which have been reconstituted from lyophilized
powder form for storage at 4°C should be centrifuged according to the
manufacturer’s instructions or locally documented procedures to
remove microaggregates prior to use in preparation of working stains. Q4.000
HLA TYPING BY FLOW CYTOMETRY (e.g. HLA B27) Q4.100
Terminology used must be defined and/or conform to nomenclature
recommended/approved by the most recent W.H.O. nomenclature committee
meeting. Q4.200
Cell Preparation. Q4.210
The method used for cell preparation should be documented to yield
appropriate preparations of viable cells. Q4.220
The viability of cell preparations should be recorded and should
exceed the laboratory’s established minimum standards for each
procedure used. Q4.2300
Labelling of specimens. Q4.2310
A negative reagent control(s) shall be run for each test cell
preparation. This control should consist of monoclonal antibody(ies)
of the same species and subclass and should be prepared/purified in
the same way as the monoclonal(s) used for phenotyping. Negative
reagent controls should consist of: Q4.2311
For indirect labelling, an irrelevant primary antibody, if available,
and in all cases, the same secondary antibody(ies) conjugated with the
same fluorochrome(s) used in all relevant test combinations. Q4.2312
For direct labelling, an irrelevant antibody conjugated with the same
fluorochrome and at the same fluorochrome: protein ratio used in all
relevant test combinations. Q4.2320
Whether analyzed directly or fixed prior to analysis, labelled cells
must be analyzed within a time period demonstrated by the laboratory
to avoid significant change in test results. Control samples must be
analyzed within the same period after staining as the test samples. Q4.3000
Reagents. Q4.3100
The specificity of monoclonal antibodies shall be verified through
tests with appropriate control cells prepared and tested by the same
method employed in the laboratory’s test sample. Q4.3200
Cell controls must be tested for each batch of monoclonal antibodies
received. Q4.3210
The control cells must include at least five cells known to express
the specified antigen. Q4.3220
The control cells must also include two cells for each crossreacting
antigen which might be confused with the specific antigen. Q4.3230
The control cells must also include at least two cells lacking the
specific and crossreacting antigens. Q4.3300
The quantities of reagents used for each test sample must be
determined by the manufacturers or from published data and whenever
possible should be verified locally by appropriate titration
procedures. Q4.3400
Reagents must be stored according to manufacturer’s instructions or
according to conditions verified to maintain stability by documented
local tests. Q4.3500
Monoclonal antibodies which have been reconstituted from lyophilized
powder form for storage at 4 degrees centigrade should be centrifuged
according to the manufacturer’s instructions or locally documented
procedures to remove microaggregates prior to use in preparation of
working stains. Q4.3600
A single monoclonal antibody may be used to define an antigen provided
its monospecificity has been sufficiently verified by local testing. Q4.3700
Minimum reactivity for assignment of a positive reaction must be
established by the laboratory. Q4.3800
If the monoclonal antibody(ies) is (are) known or found to react with
antigens other than the one specified, a written protocol must explain
how its presence or absence is finally determined. |
Board
of Directors
|